NEUROPLAY-8M

Wireless EEG system
Description
Specifications
Additionally
Downloads
FAQ

Eight-channel wireless EEG system.

It includes an elastic cap with 10 Ag / AgCl electrodes located according to the 10-20 scheme and a wireless amplifier. The system can be used for research purposes in studying brain rhythms, as well as for recording and analyzing data in educational and research institutions.

The system is used to control brain-computer interfaces (BCI) and also for neurobiological feedback (BIOS), in neuromarketing, for neurohacking, for neuroyoga, 
for neuroeducation, for neurosport, including cybersport and etc.

  • Number of EEG recording channels - 8 + GND RF
  • Monopolar
  • Frequency range -  at -3 dB level - from 1 to 50 Hz
  • Sample frequency - Hz 125 Hz, may increase to 945 Hz (when using a single channel registration)
  • Input voltage range - µV ± 300
  • Bluetooth 4.0
  • Work range from PC Up to 15 meters
  • 500 mA battery
  • Work duration - 24 hours
  • Standby time - 150 days
  • Amplifier weight - 40 grams
  • Dimensions of the amplifier, 35x55x15 mm
  • Windows, Mac IOS, Android, IPhone.

Software Cortex

Требование к программе регистрации и анализа 

  • Операционная система Windows 10 c Microsoft .NET Framework 3.5
Installer Cortex 1.9.15 Windows 10 (32/64) 103.37 Мб 2021-07-21
Основы ЭЭГ.pdf Все 6,45 Мб 2020-08-18

NeuroPlay connection

NeuroPlay does not connect to computer

If the LED on the back of the device flashes red, you need to put the device on charge.

If the LED blinks green frequently while data acquisition is stopped or disconnected from the computer, restart the amplifier by connecting it to the USB connector for two seconds.

The device while it is charging cannot be connected to the computer. You must disconnect it from the power supply.

If none of the above helps, shake the device, it should activate.

If the device is successfully activated, the LED will start blinking green once every three seconds.

NeuroPlay connects to the program, but after 2-3 seconds Cortex software closes

Make sure the amplifier you are using has not been connected to another computer.

Make sure that no more than one device is connected on one computer. 

NeuroPlay connected to computer but not connecting to Cortex software

Possible reasons

Solution

Windows 10 OS

OS must be installed at least Windows 10

You haven't updated Windows 10 for a long time

You need to install updates and restart your PC

Outdated Bluetooth drivers

Update Bluetooth drivers

The amplifier you are using is connected to another computer

Disconnect from the other PC and reconnect to the one you intend to work on
More than one device is connected to the computer Go to Settings, Bluetooth section. Disconnect other NeuroPlay devices

 

Cortex software problems

The signal does not pass (all channels display straight, flat signal lines)

Make sure the cap or headset is well-worn (fits snugly, all electrodes fit well to the scalp)

Make sure you are wearing the Rf and GND ear electrodes.

Adjust the zoom to the optimum for displaying 30 mm/s 10 µV/mm.

Poor signal quality, noise in the program window.

Make sure you are wearing the Rf and GND ear electrodes.

Adjust the electrodes on the head, ensure good contact with the scalp, try to remove hair from the places where the electrodes come into contact with the skin.

Direct signal line on one or more channels

Adjust the electrodes on the head, ensure good contact with the scalp, try to remove hair from the places where the electrodes come into contact with the skin.

Check cable for breaks by shorting the problem electrode with Rf and with GND at the same time. If the signal line is straight, contact the manufacturer to replace the electrode.